NIST’s New Scanning Probe Microscope is Supercool

Now, pushing the regime of the very cold into the very small, a research team from the National Institute of Standards and Technology (NIST), the University of Maryland, Janis Research Company, Inc., and Seoul National University, has designed and built the most advanced ultra-low temperature scanning probe microscope (ULTSPM) in the world.

Detailed in a recent paper,* the ULTSPM operates at lower temperatures and higher magnetic fields than any other similar microscope, capabilities that enable the device to resolve energy levels separated by as small as 1 millionth of an electron volt. This extraordinary resolution has already resulted in the discovery of new physics (see “Puzzling New Physics from Graphene Quartet's Quantum Harmonies”).

“To get these kinds of measurements, you need to combine coarse and extremely fine movement (the mechanical positioning of a probe tip about two atoms' distance from the sample surface), ultra-high vacuum, cryogenics and vibration isolation,” says NIST Fellow Joseph Stroscio, one of the device's co-creators. “We designed this instrument to achieve superlative levels of performance, which, in turn, requires achieving nearly the ultimate in environmental control.”

The NIST team had to overcome many technical challenges to achieve this level of precision and sensitivity, according to Young Jae Song, a postdoctoral researcher who helped develop the instrument at NIST.

Past designs used mechanical systems to move the probe tip that did not work over a wide range of temperatures. Researchers overcame this by creating piezoelectric actuators that expand with atomic scale precision when voltage is applied.

For vibration control, the group built the ULTSPM facility on top of a separate 110-ton concrete block buffered by six computer-controlled air springs. The ULTSPM, itself, sits on a 6-ton granite table, isolated from the concrete block by another set of computer-controlled air springs.

To achieve the ULTSPM's ultra low operating temperature of 10 millikelvins, the team designed a low noise dilution refrigerator to supplement the device's chilly 3-meter deep, 250-liter liquid helium bath. Because electromagnetic radiation entering through wires and cables can heat up the microscope, the ULTSPM lab is nested inside a separate, electromagnetically shielded room.

In order to ready new samples and probes without disturbing ongoing measurements, experimenters built a vacuum-sealed “railroad” system that they can disconnect from the chamber.

“The ability to create these kinds of experimental conditions opens up a whole new frontier in nanoscale physics,” says Robert Celotta, founding director of the NIST Center for Nanoscale Science and Technology. “This instrument has been five years in the making, and we can't help but be excited about all the discoveries waiting to be made.”

* Y. Song, A. Otte, V. Shvarts, Z. Zhao, Y. Kuk, S. Blankenship, A. Band, F. Hess and J. Stroscio. A 10 mK scanning probe microscopy facility. Review of Scientific Instruments. In press.

Media Contact

Mark Esser Newswise Science News

More Information:

http://www.nist.gov

All latest news from the category: Physics and Astronomy

This area deals with the fundamental laws and building blocks of nature and how they interact, the properties and the behavior of matter, and research into space and time and their structures.

innovations-report provides in-depth reports and articles on subjects such as astrophysics, laser technologies, nuclear, quantum, particle and solid-state physics, nanotechnologies, planetary research and findings (Mars, Venus) and developments related to the Hubble Telescope.

Back to home

Comments (0)

Write a comment

Newest articles

Webb captures top of iconic horsehead nebula in unprecedented detail

NASA’s James Webb Space Telescope has captured the sharpest infrared images to date of a zoomed-in portion of one of the most distinctive objects in our skies, the Horsehead Nebula….

Cost-effective, high-capacity, and cyclable lithium-ion battery cathodes

Charge-recharge cycling of lithium-superrich iron oxide, a cost-effective and high-capacity cathode for new-generation lithium-ion batteries, can be greatly improved by doping with readily available mineral elements. The energy capacity and…

Novel genetic plant regeneration approach

…without the application of phytohormones. Researchers develop a novel plant regeneration approach by modulating the expression of genes that control plant cell differentiation.  For ages now, plants have been the…

Partners & Sponsors